IVIS
®
Lumina XRMS Series III Hardware Manual
Appendix C | Scintillation Plate Holder
83
C.2 Replacing the Scintillation Plate Holder
1.
Insert the two pins on the scintillation plate holder into the L-bracket (
).
2.
Tighten the 9/16" Hex screw.
3.
Move the scintillation plate holder under the scintillator cover by horizontally rotating the plate
holder 90 degrees clockwise (
4.
Close the imaging chamber door and turn on the power to the Lumina XRMS at the rear main
power switch.
5.
Re-initialize the instrument through the Living Image
®
software.
This moves the plate holder to the home position by locating the limit sensors.
Figure C.3
Insert Scintillation Plate Holder on L-Bracket
Figure C.4
Move the Scintillation Plate Holder Underneath the Scintillation Cover
Pins on scintillation plate holder