IVIS
®
Lumina XRMS Series III Hardware Manual
Chapter 2 | Important Safety Instructions
5
2.5 Environmental Considerations for the System Components
Location for the IVIS Lumina XRMS Series III
Before the IVIS Lumina XRMS Series III is installed, consider the proper environment for the
components. Install the equipment in an environment where:
The temperature does not fluctuate widely and is maintained between 15-25 °C (59-77 °F).
The humidity does not exceed 80%.
No strong electric or magnetic fields exist.
No vibrations are present.
No corrosive gases are present.
High amounts of dust are not present.
No open flame is present.
There is sufficient space behind the IVIS Lumina XRMS Series III equipment. A minimum space
of four inches from the flat surface of the rear panel should be provided behind the IVIS Lumina
XRMS Series III to provide unobstructed air flow and access to the main power on/off switch.
The work space is level.
Heat
The system should be situated away from heat sources such as open flames, radiators, heat registers,
stoves, and other heat-generating electrical equipment.
Water and Moisture
2.6 Cleaning or Moving the System Components
Cleaning/Liquid Entry
VOLTAGE!
Do not use this product near water (for example, near a sink or wet
room) due to risk of electric shock, electrical damage, and/or equipment failure.
VOLTAGE!
Do not use liquid or aerosol cleaners and never spill liquid of any kind
on any of the IVIS Lumina XRMS Series III components. Sprays and liquids that
come into contact with the IVIS Lumina XRMS Series III hardware may result in
damage to the system or electrocution. For more details on proper care of the
system, see