IVIS
®
Lumina XRMS Series III Hardware Manual
Appendix A | XWS-260 Workstation
78
A.1 Shutting Down the Imaging System
1.
Close the Living Image
®
software and save any information at the prompt.
2.
Return the X-ray ON key selector switch to the OFF position and remove the key.
3.
Turn off the IVIS Lumina XRMS Series III imaging chamber.
4.
Turn off the computer using the standard Windows
®
shut down procedure.
5.
Turn off the power to the other system components and power surge protection devices.
6.
Unplug the devices from the wall.
If you have any problems during the shut down procedure, please contact PerkinElmer Technical
Support for assistance (see
A.2 Moving the Imaging System on the XWS-260 Workstation
1.
Unlock the wheels on the XWS-260 workstation and carefully roll the workstation to the new
location.
2.
Lock the wheels when you are finished moving the workstation.
Verify that the power outlets at the new location meet the power requirements for the system. See
for equipment power requirements,.
A.3 Starting the Imaging System
1.
Plug the devices into the wall sockets in the new location.
2.
Turn on the power surge protection devices.
3.
Turn on the computer and monitor.
4.
Turn on the IVIS Lumina XRMS Series III imaging chamber (the power switch is located on the
back of the unit) and verify that the other components such as the camera power supply, X-ray
controller (both on the back of the unit), and fluorescence lamp are also turned to the On position.
5.
Start the Living Image
®
software after the desktop screen is displayed.
6.
Enter a User ID (up to three letters) when prompted, then click Done.
7.
To initialize the system, click Initialize IVIS System in the IVIS Acquisition Control panel
(
NOTE:
The Emergency OFF switch is not intended as a main X-ray source control and should not
be used to turn the X-ray function ON or OFF on a routine basis. It should only be used in the
unlikely situation where the X-ray source must be immediately turned OFF. Under normal
circumstances, it should be left in the ON position and left as is.
!
CAUTION:
When moving the imaging system on the XWS-260 workstation, be
sure to grasp and push the workstation from as low as comfortably possible.
PerkinElmer recommends having two people present during any major move
to minimize the risk of damage to the imaging system. If the IVIS Lumina
XRMS Series III is damaged during movement, PerkinElmer cannot be held
responsible and any warranty will be voided.