Chapter 5 | X-ray Safety and Radiation Hazards
20
Turning Off X-rays
The normal procedure for turning off the x-ray source is to either let Living Image software finish
the imaging session and turn the source off as part of the normal termination of the imaging sequence,
or by terminating the program by stopping the session by "pushing" the computer stop button as
shown in Figure 4. X-rays will be stopped if the Key switch is turned OFF or if the emergency OFF
switch is pushed (see Figure 5). Finally, any attempt to open the door will activate the primary
interlock thus turning off the X-rays. Abnormal turning off of the x-rays will require re-arming the
yellow push button switch on the main console control panel.
Figure 5.4
Control Panel – Living Image Software
Figure 5.5
Emergency OFF Switch
Click to terminate
X-rays
Push in to turn off.
Turn counterclockwise to reset.