IVIS
®
Lumina XRMS Series III Hardware Manual
Chapter 8 | Fluorescence Module
57
8.2 Installation Requirements
The fluorescence equipment requires 90 - 260 VAC 50/60 Hz electrical power. The system
automatically accepts the required voltage.
8.3 Specifications
Electrical Power and Fuses
Environmental
Lamp and Fuse
Ventilation Requirements
Provide sufficient space (minimum 6 inches or 15 cm) behind the fan of the light source module so
that airflow is unobstructed. Provide a similar minimum distance behind the IVIS Lumina XRMS
Series III to enable fan cooling as well as adequate room for fiber optic cable routing.
IMPORTANT:
The fluorescence equipment operates at the same voltage as the IVIS
Lumina XRMS Series III chamber and must not be used at other than its labeled voltage.
Voltage Available
90 – 260 VAC, 50/60 Hz
Power Consumption
150
W
NOTE:
See
for more information on additional electrical power
requirements of the IVIS Lumina XRMS Series III.
Temperature
15-25
°
C
Humidity
0% to 80% Non-Condensing
Type of Use
Indoor (Pollution Degree 2)
Sound Level
61 dB <500 Hz
Altitude Ratin
<2000 meters (6560 feet)
Lamp
150 W tungsten halogen EKE bulb with an Extended Range (ER)
coating to for increased brightness in the near infrared (NIR).
Lamp voltage: 20V
Color temperature: 2856 K
°
Fuse
2.0 A, 5x20 mm, 250 VAC Slow Blow