IVIS
®
Lumina XRMS Series III Hardware Manual
Chapter 8 | Fluorescence Module
66
6.
Examine the replacement spare fuses and verify that they are the correct rating: Fuse value: 2.0A,
5 x 20, 250V SLOW BLOW.
7.
Remove fuse cassette and remove the two blown fuses (
). If only one is blown, it may
be a good idea to replace both. Place the replacement fuses into the fuse cassette. The fuses will
work in either direction.
8.
Push the fuse cassette into the fuse holder and shut the fuse door. The door should click shut.
9.
Reattach the AC line cord to the power entry module, then to the AC receptacle on the surge
protector.
10.
Return the ON/OFF switch to the ON(1) position (the red indicator will be visible on the switch).
11.
Turn on the main power switch at the rear of the IVIS Lumina XRMS Series III and start the
Living Image software.
12.
Resume normal operation.
Lamp Replacement
Contact PerkinElmer Technical Service (see
) for lamp replacement. Replacing the lamp
requires re-calibration of the IVIS Lumina XRMS Series III. The procedure also requires removing
the lamp module cover, which potentially exposes the user to hazardous voltages. Removing the lamp
module cover will void the warranty.
Figure 8.11
Fluorescent Light Source Module – Fuse Holder and Fuse Cassette
NOTE:
The power switch on the back of the light source module (
) must be in the
ON position in order for Living Image software to control lamp functions.
ON/OFF power switch
Fuse cassette
Fuse holder