IVIS
®
Lumina XRMS Series III Hardware Manual
Chapter 2 | Important Safety Instructions
6
Moving the IVIS Lumina XRMS Series III
You can move the light source module on the laboratory bench within the extent of the fiber optic
cable. Be careful not to bend the fiber optic cable (minimum radius of bending: 3 inches/7.6 cm).
2.7 Power Considerations
Power Sources
The IVIS Lumina XRMS Series III is configured for the voltage requirements of the installation
locality that was specified at the time of order. If the IVIS Lumina XRMS Series III is moved to
another area, make sure that the same voltage requirements exist.
Power Cord Protection
Power supply cords should be routed so that they are unlikely to be walked on or pinched by items
placed upon or against them. Pay close attention to receptacles and to points of connection between
cords and equipment.
Lightning and Power Line Surges
The IVIS Lumina XRMS Series III is supplied with a surge protector. All components should be
connected to this device to protect against electrical transient events. Failure to isolate the camera
from electrical transients may result in damage to the CCD camera.
Power Outages
If the IVIS Lumina XRMS Series III experiences a loss of supply power, turn off the power switch
for all components and do not restart the system until reliable power has been restored.
!
CAUTION:
The IVIS Lumina XRMS Series III is sensitive, scientific equipment and
should not be moved by any user unless the system is located on the XWS-260
workstation. Due to the risk of potential damage, it is critical that only a trained
PerkinElmer technician moves the IVIS Lumina XRMS Series III. If it is necessary
to relocate the instrument, contact PerkinElmer Technical Support (see
).
See
for more details on moving the IVIS Lumina XRMS
Series III on the XWS-260 workstation.
VOLTAGE!
The IVIS Lumina XRMS Series III can operate at multiple voltages (100-
240 VAC); however, you are not permitted to change the input voltage to any of
the system components. Several internal modifications are required for voltage
change. If the operating voltage must be changed, contact PerkinElmer Technical
Support (see