Appendix C Scintillation Plate Holder
Removing the Scintillation Plate Holder
Replacing the Scintillation Plate Holder
There may be occasions when the scintillation plate holder has to be removed for replacement or for
a more thorough cleaning than can be accomplished in situ. This appendix explains how to remove
the holder.
For this procedure, you will need the 9/64" Hex screw driver from the Scintillation Maintenance Kit
(PN 126373) or a looped T-handle Hex key (CLS138501) that was provided with the IVIS Lumina
XRMS Series III Imaging System. This kit also contains some additional screws in case some are lost
when conducting this procedure. If screws are lost inside the imaging chamber, you should try to
recover them to prevent damage to the mechanisms under the stage.
C.1 Removing the Scintillation Plate Holder
1.
Close the Living Image
®
software and turn off the power to the IVIS Lumina XRMS Series III at
the main power switch that is located at the rear of the instrument near the power cord.
2.
Disconnect the manifold gas hose on the right hand side of the stage and move the manifold out
of the way. The gas hose can be removed from the barbed fitting on the stage or from the right
hand barbed elbow on the manifold (
).
Figure C.1
Gas Manifold Connections
Gas tubing connected at
the right-hand elbow on
the gas manifold
Gas tubing connected at
the barbed fitting on stage
Manifold gas hose disconnected from the right-hand
elbow