Maintenance . 151
Press the
ENT
key.
v.
The 1st monitor screen appears.
vi.
(vi) In this status, place a paper strip at the Ex beam exit center of the
sample compartment as shown in the following figure and observe the
condition of the excitation beam.
Repeat the VERTICAL, HORIZONTAL and FOCUS adjustments in this
order little by little until the excitation beam (rainbow color) becomes
brightest. The coarse adjustment of the light source position is thus
completed. Please adjust VERTICAL first in installation because the lamp
position is shifted downward for fixation of the transport screw.
Fig. 6-21 Observation of Ex Beam
b. Fine adjustment
Following the coarse adjustment, set the excitation (Ex) wavelength at 350 nm
make a fine adjustment while observing the data. Take the procedure given below
(see chapter 5, Operation).
i. Re-mount the flow cell unit.
Observe excitation
beam on paper strip
Summary of Contents for 200A Series
Page 1: ...SERIES 200A FLUORESCENCE DETECTOR User s Guide LIQUID CHROMATOGRAPHY ...
Page 5: ...Introduction ...
Page 6: ......
Page 15: ...Safety Practices ...
Page 16: ......
Page 17: ......
Page 33: ...Safety Practices 33 Figure 2 2 Overview of the warning labels on the front of the instrument ...
Page 35: ...System Description 35 System Description ...
Page 36: ...36 Series 200a Fi User s Guide ...
Page 55: ...Installation 55 Installation ...
Page 56: ...56 Series 200a Fi User s Guide ...
Page 66: ...66 Series 200a Fi User s Guide ...
Page 67: ...Operation 67 Operation ...
Page 68: ...68 Series 200a Fi User s Guide ...
Page 117: ...Maintenance 117 Maintenance ...
Page 118: ...118 Series 200a Fl User s Guide ...
Page 154: ...154 Series 200a Fl User s Guide ...
Page 155: ...Troubleshooting ...
Page 156: ...156 ...
Page 162: ...162 Series 200a Fl User s Guide ...
Page 163: ...Appendix 163 Appendix ...
Page 164: ...164 Series 200a Fl User s Guide ...
Page 170: ...170 Series 200a Fl User s Guide ...