Operation
101
Press numeric keys 2 5 0 to 9 0 0 and then ENT key.
Setting range: 250 to 900 nm
It is recommended to set the same wavelength as was set for Em of the measuring
wavelengths.
Initial (default) value: blank
The cursor next moves to BASE.
IV.
Set the BASE (baseline processing).
<Step setting screen (BASE indicated)>
TIME Ex Em BASE PMT
0.0 250 350 1
1
Press a numeric key (1 or 2; usually 1) and then ENT key.
1: AUTO ZERO is executed. Data baseline is set to the zero point of the
recorder.
2: HOLD is executed. Baseline is held so that it doesn’t fluctuate. HOLD function
is released by executing AUTO ZERO.
Initial (default) value: 1
V.
Set the PMT (photomultiplier voltage).
<Step setting screen (PMT indicated)>
TIME Ex Em BASE PMT
0.0 250
350 1
3
1
Press a numeric key 1 to 5 and then ENT key.
VI.
When the setting for the first step is completed, return to the TIME setting on
step setting screen.
If some parameters will be changed, advance to the second and subsequent steps.
If parameters will not be changed, advance to the final step setting.
e.
Setting of second and subsequent steps of time program:
i.
Set the TIME to be changed.
<Step setting screen (TIME indicated)>
TIME Ex Em BASE PMT
_
10.0
1
Summary of Contents for 200A Series
Page 1: ...SERIES 200A FLUORESCENCE DETECTOR User s Guide LIQUID CHROMATOGRAPHY ...
Page 5: ...Introduction ...
Page 6: ......
Page 15: ...Safety Practices ...
Page 16: ......
Page 17: ......
Page 33: ...Safety Practices 33 Figure 2 2 Overview of the warning labels on the front of the instrument ...
Page 35: ...System Description 35 System Description ...
Page 36: ...36 Series 200a Fi User s Guide ...
Page 55: ...Installation 55 Installation ...
Page 56: ...56 Series 200a Fi User s Guide ...
Page 66: ...66 Series 200a Fi User s Guide ...
Page 67: ...Operation 67 Operation ...
Page 68: ...68 Series 200a Fi User s Guide ...
Page 117: ...Maintenance 117 Maintenance ...
Page 118: ...118 Series 200a Fl User s Guide ...
Page 154: ...154 Series 200a Fl User s Guide ...
Page 155: ...Troubleshooting ...
Page 156: ...156 ...
Page 162: ...162 Series 200a Fl User s Guide ...
Page 163: ...Appendix 163 Appendix ...
Page 164: ...164 Series 200a Fl User s Guide ...
Page 170: ...170 Series 200a Fl User s Guide ...