160 Series 200a Fl User’s Guide
Error Messages
Message Description
Remedy
ROM ERROR
PRESS “CL” KEY TO CLEAR
MESSAGE.
ROM is faulty.
1) Press CL key, and 1st
monitor screen returns.
2) Contact our nearest
service center.
RAM ERROR PRESS “CL” KEY
TO CLEAR MESSAGE.
RAM is faulty.
1) Press CL key, and 1st
monitor screen returns.
2) Contact our nearest
service center.
PARAMETER ERROR
PARAMETERS ARE SET TO
DEFAULT VALUE.
Analytical parameters are
inadequate.
1) Press CL key, and 1st
monitor screen returns.
2) Turn on power again.
3) Contact our nearest
service center.
LOG INFORMATION ERROR
LOG INFORMATION IS
INITIALIZED.
Log information is
inadequate.
1) Press CL key, and 1st
monitor screen returns.
2) Turn on power again.
3) Contact our nearest
service center.
EPROM ERROR INSTRUMENT
INFO. IS INITIALIZED.
Instrument parameters are
inadequate.
1) Press CL key, and 1st
monitor screen returns.
2) Turn on power again.
3) Contact our nearest
service center.
Ex (Em) SIDE OF WL DRIVE
MECHANISM ERROR
Failure in detecting Ex or Em
wavelength origin during
self-diagnosis test
(initialization).
1) Cannot be released by key
operation.
2) Turn on power again.
3) Contact our nearest
service center.
Ex (Em) SIDE ACCURACY
(254 nm) NOT DETECTED
Wavelength error exceeds
(10 nm during wavelength
accuracy check.
Contact our nearest service
center.
LAMP OFF BY OVERHEAT!
PRESS “CL” KEY TO CLEAR
MESSAGE.
Abnormal temperature
detected.
1) Press CL key, and 1st
monitor screen returns.
2) Contact our nearest
service center.
Xe lamp turned off to
prevent its overheating.
1) Press CL key, and 1st
monitor screen returns.
2) Use UTLTY key to light up
the lamp again.
LAMP OFF BY EXTERNAL
INPUT PRESS “CL” KEY TO
TURN ON.
Lamp has been turned off
due to an external contact
input. Busy status is
assumed.
1) Release the key lock if it
has been activated.
2) Press CL key and light up
the lamp again.
LAMP OFF BY EXTERNAL
ERROR PRESS “CL” KEY TO
TURN ON.
Lamp has been turned off
due to an external error.
Busy status is assumed.
1) Release the key lock if it
has been activated.
2) Press CL key and light up
the lamp again.
LAMP OFF BY LEAKAGE
PRESS “CL” KEY TO CLEAR
MESSAGE.
Liquid leakage has occurred.
Xe and Hg lamps have been
turned off. Busy status is
assumed.
1) Press CL key, and 1st
monitor screen returns.
2) Use UTLTY key to light up
the lamp again.
Summary of Contents for 200A Series
Page 1: ...SERIES 200A FLUORESCENCE DETECTOR User s Guide LIQUID CHROMATOGRAPHY ...
Page 5: ...Introduction ...
Page 6: ......
Page 15: ...Safety Practices ...
Page 16: ......
Page 17: ......
Page 33: ...Safety Practices 33 Figure 2 2 Overview of the warning labels on the front of the instrument ...
Page 35: ...System Description 35 System Description ...
Page 36: ...36 Series 200a Fi User s Guide ...
Page 55: ...Installation 55 Installation ...
Page 56: ...56 Series 200a Fi User s Guide ...
Page 66: ...66 Series 200a Fi User s Guide ...
Page 67: ...Operation 67 Operation ...
Page 68: ...68 Series 200a Fi User s Guide ...
Page 117: ...Maintenance 117 Maintenance ...
Page 118: ...118 Series 200a Fl User s Guide ...
Page 154: ...154 Series 200a Fl User s Guide ...
Page 155: ...Troubleshooting ...
Page 156: ...156 ...
Page 162: ...162 Series 200a Fl User s Guide ...
Page 163: ...Appendix 163 Appendix ...
Page 164: ...164 Series 200a Fl User s Guide ...
Page 170: ...170 Series 200a Fl User s Guide ...