Maintenance . 123
viii.
The other setting screen appears.
Press numeric key 1 and then ENT to select Em BANDWIDTH.
ix.
The Em bandwidth setting screen appears.
Press numeric key 1 or 2 and then ENT to set an emission spectrum bandwidth.
*1:
STANDARD= emission spectrum bandwidth 15 nm
2:
WIDE=emission spectrum bandwidth 30 nm.
x.
The 1
st
monitor screen returns.
d.
Setting of spectrum measurement:
i.
Press the SPECT key. The scan mode setting screen appears.
Press numeric key 2 and then ENT to select FL (Em)
ii.
Set a scanning range.
1.
Enter a starting wavelength (minimum value) for WL1.
Press 3 5 0 and ENT to set WL1.
2.
Enter an end wavelength (maximum value) for WL2.
Summary of Contents for 200A Series
Page 1: ...SERIES 200A FLUORESCENCE DETECTOR User s Guide LIQUID CHROMATOGRAPHY ...
Page 5: ...Introduction ...
Page 6: ......
Page 15: ...Safety Practices ...
Page 16: ......
Page 17: ......
Page 33: ...Safety Practices 33 Figure 2 2 Overview of the warning labels on the front of the instrument ...
Page 35: ...System Description 35 System Description ...
Page 36: ...36 Series 200a Fi User s Guide ...
Page 55: ...Installation 55 Installation ...
Page 56: ...56 Series 200a Fi User s Guide ...
Page 66: ...66 Series 200a Fi User s Guide ...
Page 67: ...Operation 67 Operation ...
Page 68: ...68 Series 200a Fi User s Guide ...
Page 117: ...Maintenance 117 Maintenance ...
Page 118: ...118 Series 200a Fl User s Guide ...
Page 154: ...154 Series 200a Fl User s Guide ...
Page 155: ...Troubleshooting ...
Page 156: ...156 ...
Page 162: ...162 Series 200a Fl User s Guide ...
Page 163: ...Appendix 163 Appendix ...
Page 164: ...164 Series 200a Fl User s Guide ...
Page 170: ...170 Series 200a Fl User s Guide ...