122.
Series 200a Fl User’s Guide
iii.
The 1st monitor screen appears again.
(c)
Set a time constant and a photomultiplier voltage.
i.
Press the UTLTY key, and the utility setting screen appears.
Press numeric key 2 and then ENT to select time constant.
ii.
The time constant setting screen appears.
Press numeric key 5 to select 2.0 s and then ENT.
iii.
The 1st monitor screen appears.
iv.
Press UTLTY key and the utility setting screen reappears.
Press numeric key 4 and then ENT to select photomultiplier voltage.
v.
The pmt voltage setting screen appears.
Press numeric key 3 and then ENT to select MID.
vi.
The 1st monitor screen reappears.
vii.
Press UTLTY key, and the utility setting screen reappears.
Press numeric key 5 and then ENT to select OTHERS.
Summary of Contents for 200A Series
Page 1: ...SERIES 200A FLUORESCENCE DETECTOR User s Guide LIQUID CHROMATOGRAPHY ...
Page 5: ...Introduction ...
Page 6: ......
Page 15: ...Safety Practices ...
Page 16: ......
Page 17: ......
Page 33: ...Safety Practices 33 Figure 2 2 Overview of the warning labels on the front of the instrument ...
Page 35: ...System Description 35 System Description ...
Page 36: ...36 Series 200a Fi User s Guide ...
Page 55: ...Installation 55 Installation ...
Page 56: ...56 Series 200a Fi User s Guide ...
Page 66: ...66 Series 200a Fi User s Guide ...
Page 67: ...Operation 67 Operation ...
Page 68: ...68 Series 200a Fi User s Guide ...
Page 117: ...Maintenance 117 Maintenance ...
Page 118: ...118 Series 200a Fl User s Guide ...
Page 154: ...154 Series 200a Fl User s Guide ...
Page 155: ...Troubleshooting ...
Page 156: ...156 ...
Page 162: ...162 Series 200a Fl User s Guide ...
Page 163: ...Appendix 163 Appendix ...
Page 164: ...164 Series 200a Fl User s Guide ...
Page 170: ...170 Series 200a Fl User s Guide ...