152.
Series 200a Fl User’s Guide
ii. Press
the
WL
key and the wavelength setting screen appears. Set an
excitation (Ex) wavelength.
Press numeric keys
3 5 0
and then
ENT
to set the Ex.
iii.
Emission (Em) wavelength will not be set here. Leave the present setting
as it is.
Press the
ENT
key.
iv.
The 1st monitor screen appears.
v.
Press
or
key and the 2nd monitor screen appears.
vi.
In the present status, turn the FOCUS, HORIZONTAL and VERTICAL
adjusting screws gradually in a repetitive manner while watching the
MONITOR (Ex) value on the screen until this value is maximum. When a
maximum value is reached, the fine adjustment of the lamp position is
completed.
vii.
Press or
ESC
key and the 1st monitor screen will return.
4.
Entry of Lamp Replacement Date
When the above procedure is finished, enter a lamp replacement date. With this entry, the
cumulative on-time is set to “0” and the number of lightings to “1”.
Summary of Contents for 200A Series
Page 1: ...SERIES 200A FLUORESCENCE DETECTOR User s Guide LIQUID CHROMATOGRAPHY ...
Page 5: ...Introduction ...
Page 6: ......
Page 15: ...Safety Practices ...
Page 16: ......
Page 17: ......
Page 33: ...Safety Practices 33 Figure 2 2 Overview of the warning labels on the front of the instrument ...
Page 35: ...System Description 35 System Description ...
Page 36: ...36 Series 200a Fi User s Guide ...
Page 55: ...Installation 55 Installation ...
Page 56: ...56 Series 200a Fi User s Guide ...
Page 66: ...66 Series 200a Fi User s Guide ...
Page 67: ...Operation 67 Operation ...
Page 68: ...68 Series 200a Fi User s Guide ...
Page 117: ...Maintenance 117 Maintenance ...
Page 118: ...118 Series 200a Fl User s Guide ...
Page 154: ...154 Series 200a Fl User s Guide ...
Page 155: ...Troubleshooting ...
Page 156: ...156 ...
Page 162: ...162 Series 200a Fl User s Guide ...
Page 163: ...Appendix 163 Appendix ...
Page 164: ...164 Series 200a Fl User s Guide ...
Page 170: ...170 Series 200a Fl User s Guide ...