8.1 Introduction
This section details the work necessary in the ‘inspection and service’ of the LinerSCAN system. It is
necessary to check the system – including, but not limited to:
•
Condition of flexible hoses
•
Condition and operation of valves
•
Condition and operation of air and power isolators
•
Checking and rectification of blockages
•
Cleaning the sensor
•
Checking calibration
•
System functionality
8.1.1 Checking the Readings – Sensor Confidence Test
A measure check can be made by inserting the calibration rod into the sensor as follows:
!
Caution:
Turn the oil valves OFF and disconnect the air supply.
Step 1:
Step 2:
Step3:
Step 4:
Remove the oil pipes
Inert check standard
rods into ‘Oil In’ and ‘Oil
Out’
Start a measure: ‘single
Measurement’ specific
to the cylinder/sensor or
‘m’ via RS232
Remove
both
rods
quickly
when
the
solenoid ‘clicks’
The result should be between the values stated on the Check Standard Rod ± 30ppm.