MA-K17510-KW LinerSCAN manual ISSUE P.doc
Page 85 of 93
8.1.2 Calibrating the Sensor
Calibration is only possible via RS232 connection. The sensors require the LinerSCAN software to be
closed – or the ‘Network Dongle’ to be removed. Follow the steps below:
Step 1:
Step 2:
Step3:
Step 4:
Remove the oil pipes
Press
‘x’
in
HyperTerminal
Press ‘c’ and follow on
screen instructions
When requested ‘Insert
Standard’
The standard can be
removed
once
calibration
data
appears
(scaling
factor).
Follow
‘confidence
check’
to
verify
calibration