7.3.1 Flat Traces and Filling Times
The sensor measurement chamber will fill at different rates. This is affected by parameters such as
engine load, fuel sulphur and cylinder oil feedrate. In some circumstances, measurements may not
occur for several hours. This is seen as flat lines on data charts:
The progress of filling can be seen by observing the level sensor within the sensor box. When the
sensor is full the LED will switch on, when empty the LED will be off. When a blast occurs the system
should wempty of oil, switching off the LED.
Should the sensor not fill [LED not illuminating] after an adequate amount of time (30 mins to 1 hour),
conduct an oil flow test. If filling does not start, pipe work inspection is recommended.