7.1 Software Problems
SYMPTOM
POSSIBLE CAUSES
SOLUTION
Software Installation does
not
work
properly/incomplete
installation
Virus checker not allowing
proper access.
Ensure virus checker is disabled and re-
install.
LinerSCAN
software
‘hangs’ or is slow
Alarm values have just
been
entered
and
Software is transmitting
data
Allow time for the sensor status to update
before clicking further buttons. The system
can take several minutes to update all
configurations on a large (many sensor)
network.
Software does not fit on the
screen
screen resolution too low
Screen resolution needs to be 1200*700
min
Computer is slow to re-start
after LinerSCAN software
install
Virus checker was ‘on’
during
installation.
LinerSCAN not installed
properly
Turn the virus checker ‘off’ and re-install
LinerSCAN. If the PC is not connected to a
network or internet consider disabling virus
checker or removing from PC – software
should run faster.
Software does not see one
sensor
Sensor not connected to
network
Check the sensor's wiring
Software/firmware problem Restart software on PC and re-boot
sensor
Fuse blown
Check lights on bottom of sensor (inside box)
Is the fuse OK?
Sensor not communicating
Check the LinerSCAN software, network (F6)
tab for errors.
LinerSCAN software not
receiving
data/measurement
from
one sensor
Sensor not connected to
network
Check all connecting plugs/leads
Software/firmware problem Restart software on laptop and re-boot
sensor
LinerSCAN software does
not see some/any sensors
Break in network
Check all connecting plugs/leads
Check the CAN terminating connector is
fitted.
Engine running indication
not operating
Fault with ship's system
Check that the ship's system is operating
properly
Bad/broken connection to
ship's system
Check connections between the junction
box and the ship's system
Air alarm not operating
Pressure switch faulty
Check/calibrate pressure switch
Loose/bad connections
Check all relevant connections between
the pressure switch and the laptop
Air alarm keeps initiating
Pressure switch faulty
Check/calibrate pressure switch
Loose/bad connections
Check all relevant connections between
the pressure switch and the laptop
Block Alarm keeps initiating Blocked Sensor
The sensor will need unblocking, see:
‘Maintenance and Servicing’ section.
Summary of Contents for FG-K17401-KW
Page 9: ...MA K17510 KW LinerSCAN manual ISSUE P doc Page 8 of 93 3 Installation Installation ...
Page 34: ...MA K17510 KW LinerSCAN manual ISSUE P doc Page 33 of 93 4 Commissioning Commissioning ...
Page 40: ...MA K17510 KW LinerSCAN manual ISSUE P doc Page 39 of 93 5 Operation Operation Manual ...
Page 89: ...MA K17510 KW LinerSCAN manual ISSUE P doc Page 88 of 93 ...