THE EXHAUST VOLUME. If possible, adjust exhaust volume with damper at 60° for maximum
control.
Step 4: Now, measure inflow velocity using the recommended procedure found in Table 7.0. If
necessary, adjust the exhaust system and/or damper to achieve the correct inflow velocity.
Use Table 7.1 to relate downflow and exhaust/inflow volumes and corresponding average
airflow velocities. It is always desirable to achieve airflow values that are set as close as possible
to nominal.
If the inflow velocity is calibrated at the outer edge of the range, a greater chance
for alarm conditions would exist due to its closer proximity to the alarm limits.
•
Once damper position and/or exhaust volume is set, press SAVE to enter the damper position.
•
Press -/ + to change exhaust volume to the value just calibrated using the inflow volume.
NOTE
: Assume downflow velocity is at the nominal value of 60 fpm for exhaust volume calculation purposes as it
will be adjusted during the downflow calibration process.
Summary of Contents for LabGard NU-560-400
Page 7: ...OM0295 Rev 2 September 2020 Page 7 of 103...
Page 9: ...OM0295 Rev 2 September 2020 Page 9 of 103...
Page 10: ...OM0295 Rev 2 September 2020 Page 10 of 103...
Page 16: ...OM0295 Rev 2 September 2020 Page 16 of 103...
Page 17: ...OM0295 Rev 2 September 2020 Page 17 of 103...
Page 32: ...OM0295 Rev 2 September 2020 Page 32 of 103 Set New Password Retype New Password...
Page 43: ...OM0295 Rev 2 September 2020 Page 43 of 103...
Page 48: ...OM0295 Rev 2 September 2020 Page 48 of 103...
Page 80: ...OM0295 Rev 2 September 2020 Page 80 of 103...
Page 96: ...OM0295 Rev 2 September 2020 Page 96 of 103...
Page 97: ...OM0295 Rev 2 September 2020 Page 97 of 103...
Page 98: ...OM0295 Rev 2 September 2020 Page 98 of 103...
Page 99: ...OM0295 Rev 2 September 2020 Page 99 of 103...
Page 100: ...OM0295 Rev 2 September 2020 Page 100 of 103...
Page 101: ...OM0295 Rev 2 September 2020 Page 101 of 103...