OM0295
Rev 2 September 2020
Page 42 of 103
Step 4: To remove the supply filter:
a. Remove plenum screws in front of supply HEPA filter.
b. Lift the permanent plenum and hold up with wire strap.
c. Carefully remove the supply filter.
DIRECT EXPOSURE SHOULD BE AVOIDED, EVEN THOUGH THE FILTER
IS IN AN UNCONTAMINATED PLENUM
.
Step 5: To install the supply filter, simply reverse the procedure outlined in the steps above.
7.3.2 Exhaust Filter Replacement
Step 1: Remove exhaust filter access panel, which is held into position by screws. Remove the screws and panel.
Step 2: Locate the external single point release bolt on the top right hand side of the cabinet.
Use a 5/16-inch (8mm) wrench to release the exhaust filter rotating counter clockwise.
Step 3: Carefully remove the exhaust filter.
Dispense of spent HEPA filters properly.
Avoid direct contact to "dirty side" of the filters.
Place in sealed bag and label waste containers/cartons based on the type of
hazard. Follow all Local, State and Federal guidelines for disposal of HEPA
filter solid waste.
When installing the new filter, apply a thin layer of silicone grease to the gasket of the filter and carefully insert
into exhaust chamber. Tighten HEPA seal frame (clockwise) until the gasket is visually depressed by 1/8 inch
(3mm). The procedure for replacing the exhaust filter with the Bag-in/Bag-out option is shown on Drawing
BCD-05660.
7.3.3 Motor/Blower Assembly Removal
a. It is recommended that the motor/blower to be removed as a single unit. To remove, disconnect
electrical connections to the motor, remove the HEPEX pressure plenum and unbolt the motor/blower
assembly from the roof of the cabinet (4 places). Always inspect the rubber isolation motor mounts and
replace those that are cracked or visibly show stress.
b. Replace the motor exactly as originally installed in the blower housing, paying particular attention to the
correct electrical connections (see Electrical Schematic).
c. Re-install the new motor/blower assembly.
7.4 Sliding Window Replacement and Manual Adjustment
The sliding window replacement is accomplished by removing the front decorative panel, control center, and window glide
assemblies. The sliding window adjustment may be required due to everyday use over the life of the cabinet. Both window
glides are adjustable by a set screw and tension screw method. When adjusting the sliding window, be sure to verify
proper micro switch operation. If the sliding window is too loose, it will not properly activate the micro switches, thus
causing potential operational malfunctions to occur.
CAUTION
!
CAUTION
!
Summary of Contents for LabGard NU-560-400
Page 7: ...OM0295 Rev 2 September 2020 Page 7 of 103...
Page 9: ...OM0295 Rev 2 September 2020 Page 9 of 103...
Page 10: ...OM0295 Rev 2 September 2020 Page 10 of 103...
Page 16: ...OM0295 Rev 2 September 2020 Page 16 of 103...
Page 17: ...OM0295 Rev 2 September 2020 Page 17 of 103...
Page 32: ...OM0295 Rev 2 September 2020 Page 32 of 103 Set New Password Retype New Password...
Page 43: ...OM0295 Rev 2 September 2020 Page 43 of 103...
Page 48: ...OM0295 Rev 2 September 2020 Page 48 of 103...
Page 80: ...OM0295 Rev 2 September 2020 Page 80 of 103...
Page 96: ...OM0295 Rev 2 September 2020 Page 96 of 103...
Page 97: ...OM0295 Rev 2 September 2020 Page 97 of 103...
Page 98: ...OM0295 Rev 2 September 2020 Page 98 of 103...
Page 99: ...OM0295 Rev 2 September 2020 Page 99 of 103...
Page 100: ...OM0295 Rev 2 September 2020 Page 100 of 103...
Page 101: ...OM0295 Rev 2 September 2020 Page 101 of 103...