OM0295
Rev 2 September 2020
Page 58 of 103
Step 3: At this point, the night setback is active and the butterfly valve will begin to close. Now measure
the inflow volume using a Direct Reading Instrument of inflow velocity measurement method.
The goal is to reduce the inflow volume to the following values, which represents approximately
100 fpm (.51m/s) inflow, or may be set at any other values within the damper operational range
of 30% to 80%.
NU-560-400
270 CFM/459 CMH
NU-560-600
410 CFM/697 CMH
Please note, in the night setback operation, the electronic airflow control system will cause the
butterfly valve actuator to hunt more operating at lower airflow volumes. When taking the
inflow volume measurements, average several readings to obtain the most accurate results.
Summary of Contents for LabGard NU-560-400
Page 7: ...OM0295 Rev 2 September 2020 Page 7 of 103...
Page 9: ...OM0295 Rev 2 September 2020 Page 9 of 103...
Page 10: ...OM0295 Rev 2 September 2020 Page 10 of 103...
Page 16: ...OM0295 Rev 2 September 2020 Page 16 of 103...
Page 17: ...OM0295 Rev 2 September 2020 Page 17 of 103...
Page 32: ...OM0295 Rev 2 September 2020 Page 32 of 103 Set New Password Retype New Password...
Page 43: ...OM0295 Rev 2 September 2020 Page 43 of 103...
Page 48: ...OM0295 Rev 2 September 2020 Page 48 of 103...
Page 80: ...OM0295 Rev 2 September 2020 Page 80 of 103...
Page 96: ...OM0295 Rev 2 September 2020 Page 96 of 103...
Page 97: ...OM0295 Rev 2 September 2020 Page 97 of 103...
Page 98: ...OM0295 Rev 2 September 2020 Page 98 of 103...
Page 99: ...OM0295 Rev 2 September 2020 Page 99 of 103...
Page 100: ...OM0295 Rev 2 September 2020 Page 100 of 103...
Page 101: ...OM0295 Rev 2 September 2020 Page 101 of 103...