OM0295
Rev 2 September 2020
Page 45 of 103
7.5.2 Calibration/Service Menu
The Calibration/Service menu provides a list of sub-menu items to accomplish all service tasks. Each sub-menu
item will be described in the following sections
For airflow calibration, only the first three sub-menu items are used in the calibration process.
7.5.2.1 Cabinet Type/ Motor Type
The cabinet type can be verified in the control system and is factory set and shouldn’t require alteration. The
cabinet type default information controls unit of measure, setpoints and limits based on the type and size of
cabinet. The OEM displayed menu item references the LOGO style displayed NuAire being the default.
To verify, press CABINET TYPE. The current type of cabinet will be designated. Again to verify, press the
correct Type and the Model/Size of the current selected cabinet model will be designated. Once verified the
correct type and model are designated, then press BACK to return to Calibration/Service menu.
Summary of Contents for LabGard NU-560-400
Page 7: ...OM0295 Rev 2 September 2020 Page 7 of 103...
Page 9: ...OM0295 Rev 2 September 2020 Page 9 of 103...
Page 10: ...OM0295 Rev 2 September 2020 Page 10 of 103...
Page 16: ...OM0295 Rev 2 September 2020 Page 16 of 103...
Page 17: ...OM0295 Rev 2 September 2020 Page 17 of 103...
Page 32: ...OM0295 Rev 2 September 2020 Page 32 of 103 Set New Password Retype New Password...
Page 43: ...OM0295 Rev 2 September 2020 Page 43 of 103...
Page 48: ...OM0295 Rev 2 September 2020 Page 48 of 103...
Page 80: ...OM0295 Rev 2 September 2020 Page 80 of 103...
Page 96: ...OM0295 Rev 2 September 2020 Page 96 of 103...
Page 97: ...OM0295 Rev 2 September 2020 Page 97 of 103...
Page 98: ...OM0295 Rev 2 September 2020 Page 98 of 103...
Page 99: ...OM0295 Rev 2 September 2020 Page 99 of 103...
Page 100: ...OM0295 Rev 2 September 2020 Page 100 of 103...
Page 101: ...OM0295 Rev 2 September 2020 Page 101 of 103...