5.4 Certification Testing Methods and Equipment
After installation and prior to use, NuAire recommends that the cabinet be certified or commissioned to factory standards.
As a part of certification, the certifier should go through the following initial checklist to assure all aspects of the LFBSC
installation are complete and ready for certification.
•
Review product installation
−
Exhaust connection
−
Damper valve installed correctly with label toward front
−
LFBSC base stand level
•
Verify airflow sensor shroud is in place
−
Downflow
•
Verify configuration type selection for specific model * (see section 7.5.2)
•
Verify setpoints and alarm limits for specific model * (see section 7.5.2)
•
Perform BCS certification
−
At a minimum, the following tests should be performed:
HEPA filter leak test
Downflow velocity test
Inflow velocity test
Airflow smoke patterns
Site installation assessment tests
•
Perform Site Assessment Tests
−
The NU-560 requires verification of the supply fan interlock and back-up pressure switch operation utilizing
independent exhaust volume measurement instrument (DIM). Per NSF/ANSI 49, a 20% loss of exhaust volume
must produce an airflow alarm within 15 seconds.
The testing methods and equipment required are specified on the factory inspection report included with this manual.
(See insert in back cover)
NOTE: IT IS RECOMMENDED THAT THESE TESTS BE PERFORMED BY A QUALIFIED TECHNICIAN WHO IS
FAMILIAR WITH THE METHODS AND PROCEDURES FOR CERTIFYING BIOSAFETY CABINETS
(SEE INSERT)
NOTE: AFTER THE INITIAL CERTIFICATION, NUAIRE RECOMMENDS THAT THE CABINET BE RECERTIFIED (AT A
MINIMUM) ON AN ANNUAL BASIS AND AFTER EVERY FILTER CHANGE OR MAINTENANCE ACTION OR
ANY TIME THE OPERATOR FEELS IT IS NECESSARY.
Note that the LABGARD® ES cabinets, filters, and seals provide premium performance. Quality Control in both design and
manufacturing assure superior reliability. However, protection to both product and operator is so vital that certification to
the performance requirements should be accomplished as stated to ensure Biosafety established by the factory standards.
* If the specific model is a special product with non-standard setpoints and alarm limits, the new values will
be located on the factory Inspection Report.
Summary of Contents for LabGard NU-560-400
Page 7: ...OM0295 Rev 2 September 2020 Page 7 of 103...
Page 9: ...OM0295 Rev 2 September 2020 Page 9 of 103...
Page 10: ...OM0295 Rev 2 September 2020 Page 10 of 103...
Page 16: ...OM0295 Rev 2 September 2020 Page 16 of 103...
Page 17: ...OM0295 Rev 2 September 2020 Page 17 of 103...
Page 32: ...OM0295 Rev 2 September 2020 Page 32 of 103 Set New Password Retype New Password...
Page 43: ...OM0295 Rev 2 September 2020 Page 43 of 103...
Page 48: ...OM0295 Rev 2 September 2020 Page 48 of 103...
Page 80: ...OM0295 Rev 2 September 2020 Page 80 of 103...
Page 96: ...OM0295 Rev 2 September 2020 Page 96 of 103...
Page 97: ...OM0295 Rev 2 September 2020 Page 97 of 103...
Page 98: ...OM0295 Rev 2 September 2020 Page 98 of 103...
Page 99: ...OM0295 Rev 2 September 2020 Page 99 of 103...
Page 100: ...OM0295 Rev 2 September 2020 Page 100 of 103...
Page 101: ...OM0295 Rev 2 September 2020 Page 101 of 103...