OM0295
Rev 2 September 2020
Page 20 of 103
LabGard® ES Energy Saver Class II, Type B2 Laminar Flow Biosafety Cabinet
Models NU-560-400/600
Catalog
Number
Catalog Number
NU-560-400
Nominal 4 foot (1.2m)
NU-560-600
Nominal 6 foot (1.8m)
Performance Specifications
1. Personal Protection
2. Product Protection
NSF/ANSI 49
NSF/ANSI 49
NSF Std. No. 49 Class
Class II, Type B2
Class II, Type B2
Style of Cabinet
Bench Top/Console w/Base Stand/
Storage Cabinet
Bench Top/Console w/Base Stand/
Storage Cabinet
Cabinet Construction
All welded stainless steel 16/18GA,
Type 304 pressure tight design
All welded stainless steel 16/18GA,
Type 304 pressure tight design
Diffuser for Air Supply (Metal)
Non-Flammable
Non-Flammable
HEPA Filter Seal Type:
Supply Filter-99.99% Eff. on 0.3 Microns
Exhaust Filter-99.99% Eff. on 0.3 Microns
HEPEX Seal
Neoprene, Spring loaded
HEPEX Seal
Neoprene, Spring loaded
Fumigation per NIH/NSF Procedure
Yes
Yes
Standard Services:
Service Coupling (3/8 inch NPT)
Gas Valve/Service Coupling (3/8 inch NPT)
Duplex Outlet
Three: Left middle, Left bottom,
Right middle Sidewalls
One, Right bottom Sidewall
One, Left Front Faring
Three: Left middle, Left bottom, Right
middle Sidewalls
One, Right bottom Sidewall
One, Left Front Faring
Optional Services: Gas Cocks 3/8" NPT
**Remote Controlled Valves
Ultraviolet Light
Up to 3 ea. Sidewall
Up to 3 ea. Sidewall
One, Backwall
Up to 3 ea. Sidewall
Up to 3 ea. Sidewall
One, Backwall
Cabinet Size Inches (mm):
Height (Fully Assembled) (Incl. Exh Filter Fastener)
Height (Minimum for Transport)
Width
Depth (with Armrest removed)
62 (1575)
62 (1575)
53 5/8 (1362)
32 3/8 (823)
62 (1575)
62 (1575)
77 5/8 (1972)
32 3/8 (823)
Work Access Opening Inches (mm):
Standard Opening Height
Standard Inflow Velocity
8 (203)
105 FPM (.53 m/s)
8 (203)
105 FPM (.53 m/s)
Work Zone Inches (mm): Height
Width
Depth measured at 8 inches (203mm)
25 3/16 (640)
46 3/8 (1178)
26 (660)
25 3/16 (640)
70 3/8 (1788)
26 (660)
Viewing Window Inches (mm):
Standard is Tempered Sliding Glass
Hinged Tempered Glass (optional)
0.375 (10 mm) Closed
18 1/2 (470) Open
8 (203) Access Opening
0.375 (10 mm) Closed
18 1/2 (470) Open
8 (203) Access Opening
Certification Exhaust Value CFM/CMH
Concurrent Balance Value CFM/CMH +
740/1257
815/1385
1122/1906
1275/2166
Plant Duct Static Pressure Eng./Metric
1.7 w.g./43mm w.g.
1.8 w.g./46mm w.g.
Heat Rejected, BTU, Per Hour
474
584
Electrical: 115V
Volts, AC (Hz)
++Amps: Blower/Lights
Amps: Outlet
Rated Amps:
12 ft. Power Cord (one)
U.L./UL-C, US Listed
115, 60
2.1
3
8
14 GA - 3 Wire, 15A
U.L./UL-C, US Listed
115, 60
2.6
3
10
14 GA-3 Wire, 15A
Crated Shipping Weight:
Net Weight
540 lbs. /245 kg.
490 lbs. /222 kg.
730 lbs. /331 kg.
680 lbs. /308 kg.
**Remote controlled valve handles project through front fairing. Decorative side panels are available to cover plumbing.
+Concurrent Balance Value shall be used for design and balance exhaust/supply HVAC requirements. Values provided are nominal. Design
tolerances should be added for measurement and system differences.
++ Based on cabinet with new filters running at 115 VAC.
Summary of Contents for LabGard NU-560-400
Page 7: ...OM0295 Rev 2 September 2020 Page 7 of 103...
Page 9: ...OM0295 Rev 2 September 2020 Page 9 of 103...
Page 10: ...OM0295 Rev 2 September 2020 Page 10 of 103...
Page 16: ...OM0295 Rev 2 September 2020 Page 16 of 103...
Page 17: ...OM0295 Rev 2 September 2020 Page 17 of 103...
Page 32: ...OM0295 Rev 2 September 2020 Page 32 of 103 Set New Password Retype New Password...
Page 43: ...OM0295 Rev 2 September 2020 Page 43 of 103...
Page 48: ...OM0295 Rev 2 September 2020 Page 48 of 103...
Page 80: ...OM0295 Rev 2 September 2020 Page 80 of 103...
Page 96: ...OM0295 Rev 2 September 2020 Page 96 of 103...
Page 97: ...OM0295 Rev 2 September 2020 Page 97 of 103...
Page 98: ...OM0295 Rev 2 September 2020 Page 98 of 103...
Page 99: ...OM0295 Rev 2 September 2020 Page 99 of 103...
Page 100: ...OM0295 Rev 2 September 2020 Page 100 of 103...
Page 101: ...OM0295 Rev 2 September 2020 Page 101 of 103...