7.10.4 Cabinet Reset
The main control board has two software operating resets available for qualified service personnel.
The two types are the following:
Factory Reset - Resets setpoints and selected option settings. Factory reset should be used in the event the
system memory develops an error in operation. Cabinet type, motor type and calibration data
will not be affected with this reset.
Master Reset - Resets all calibration, cabinet type, motor type, sensor data, and options settings back to
default settings. Master Reset should only need to be used for a main control board
replacement.
After pressing the Calibration/Service menu item, a password screen will appear. The default password is “9876”.
Once the service password is entered, the Calibration/Service menu will appear.
Select CABINET RESET from the menu.
Summary of Contents for LabGard NU-560-400
Page 7: ...OM0295 Rev 2 September 2020 Page 7 of 103...
Page 9: ...OM0295 Rev 2 September 2020 Page 9 of 103...
Page 10: ...OM0295 Rev 2 September 2020 Page 10 of 103...
Page 16: ...OM0295 Rev 2 September 2020 Page 16 of 103...
Page 17: ...OM0295 Rev 2 September 2020 Page 17 of 103...
Page 32: ...OM0295 Rev 2 September 2020 Page 32 of 103 Set New Password Retype New Password...
Page 43: ...OM0295 Rev 2 September 2020 Page 43 of 103...
Page 48: ...OM0295 Rev 2 September 2020 Page 48 of 103...
Page 80: ...OM0295 Rev 2 September 2020 Page 80 of 103...
Page 96: ...OM0295 Rev 2 September 2020 Page 96 of 103...
Page 97: ...OM0295 Rev 2 September 2020 Page 97 of 103...
Page 98: ...OM0295 Rev 2 September 2020 Page 98 of 103...
Page 99: ...OM0295 Rev 2 September 2020 Page 99 of 103...
Page 100: ...OM0295 Rev 2 September 2020 Page 100 of 103...
Page 101: ...OM0295 Rev 2 September 2020 Page 101 of 103...