OM0295
Rev 2 September 2020
Page 85 of 103
•
UV Window interlock
:
This parameter allows for the selection of the window closed switch to be
interlocked with the UV light option. When the window interlock is on, the window must be closed for
the UV light to operate. When the window interlock is off the UV light can be turned on regardless of
window position.
NOTE
: In addition to the
TOUCHLINK
system UV window interlock there is a double redundant UV window
interlock relay. To override the UV window interlock for service purposes only, both interlocks must be changed
through the
TOUCHLINK
system and shorting the relay connection (see electrical schematic for reference).
•
Blower Window Interlock
: This parameter allows for the selection of the window closed switch to be
interlocked to the blower. When the blower interlock is on, the blower will turn off when the window is
closed. When the blower is off, the blower will continue to run when the window is closed
Summary of Contents for LabGard NU-560-400
Page 7: ...OM0295 Rev 2 September 2020 Page 7 of 103...
Page 9: ...OM0295 Rev 2 September 2020 Page 9 of 103...
Page 10: ...OM0295 Rev 2 September 2020 Page 10 of 103...
Page 16: ...OM0295 Rev 2 September 2020 Page 16 of 103...
Page 17: ...OM0295 Rev 2 September 2020 Page 17 of 103...
Page 32: ...OM0295 Rev 2 September 2020 Page 32 of 103 Set New Password Retype New Password...
Page 43: ...OM0295 Rev 2 September 2020 Page 43 of 103...
Page 48: ...OM0295 Rev 2 September 2020 Page 48 of 103...
Page 80: ...OM0295 Rev 2 September 2020 Page 80 of 103...
Page 96: ...OM0295 Rev 2 September 2020 Page 96 of 103...
Page 97: ...OM0295 Rev 2 September 2020 Page 97 of 103...
Page 98: ...OM0295 Rev 2 September 2020 Page 98 of 103...
Page 99: ...OM0295 Rev 2 September 2020 Page 99 of 103...
Page 100: ...OM0295 Rev 2 September 2020 Page 100 of 103...
Page 101: ...OM0295 Rev 2 September 2020 Page 101 of 103...