OM0295
Rev 2 September 2020
Page 64 of 103
Areas/Correction Factors for Calculations
Cab.
Size
3” (76mm) Constricted Window
Access Area ft2, (m2)
8” (203mm) Window Access
Opening Area ft2, (m2)
Correction Factor for 8”
(203mm) Window
Work Zone Area
ft2, (m2)
400
.97
(.090)
2.58
(.239)
.97
7.81
(.781)
600
1.47
(.137)
3.91
(.363)
.98
11.85
(1.101)
Table 7.1
Certification Values
The following are recommended minimum/maximum cabinet airflow certification setpoints per NSF/ANSI 49. NuAire
recommends, however, operation at the stated average flow, for ease of maintenance and annual certification.
THE FOLLOWING EXHAUST FLOWS ARE FOR AN 8 INCH (203MM) WORK ACCESS OPENING:
NOTE:
1: NuAire recommends the cabinet be set up and certified at the stated
nominal average inflow
.
EXAMPLE
:
NU-560-400 Inflow Volume Measurement / Calibration
1) Inflow volume measured with DIM at cabinet face with blower on and downflow velocity approximately at 60 fpm
(.30 m/s), inflow volume = 271cfm (460 CMH).
2) Inflow supply volume = exhaust volume
271cfm (460CMH) + 469cfm (797CMH) = 740cfm (1257CMH)
3) So, based on the inflow measurement volume and adding the supply volume from the table for the nominal
average velocity, the calibrated exhaust value entered should be 740cfm (1257CMH).
NOTE:
2: For additional guidance, the values provided below with a constant 60fpm (.30mls) downflow velocity can be
used for exhaust adjustment.
NU-560-400
NU-560-600
Inflow Volume
Inflow Velocity
Exhaust Volume
Inflow Volume
Inflow Velocity
Exhaust Volume
258
100
727
391
100
1102
263
102
732
399
102
1110
268
104
737
407
104
1118
271
105
740
411
105
1122
274
106
743
415
106
1126
279
108
748
422
108
1133
284
110
753
430
110
1141
Parameter
Minimum
Stated Nominal
Maximum
Acceptable Flow
Average Flow
Acceptable Flow
NU-560-400
1. Inflow Avg. Velocity
100 (.51 m/s) FPM
105 (.53 m/s) FPM
110 (.56 m/s) FPM
2. Inflow Volume
258 (438 CMH) CFM
271 (460 CMH) CFM
284 (483 CMH) CFM
3. Supply Avg. Velocity
55 (.27 m/s) FPM
60 (.30 m/s) FPM
65 (.33 m/s) FPM
4. Supply Volume
430 (731 CMH) CFM
469 (797 CMH) CFM
508 (863 CMH) CFM
5. Total Volume
688 (1169 CMH) CFM
740 (1257 CMH) CFM
792 (1346 CMH) CFM
NU-560-600
1. Inflow Avg. Velocity
100 (.51 m/s) FPM
105 (.53 m/s) FPM
110 (.56 m/s) FPM
2. Inflow Volume
391 (664 CMH) CFM
411(698 CMH) CFM
430 (731 CMH) CFM
3. Supply Avg. Velocity
55 (.27 m/s) FPM
60 (.30 m/s) FPM
65 (.33 m/s) FPM
4. Supply Volume
652 (1108 CMH) CFM
711 (1208 CMH) CFM
770 (1308 CMH) CFM
5. Total Volume
1043 (1772 CMH) CFM
1122 (1906 CMH) CFM
1200 (2039 CMH) CFM
Summary of Contents for LabGard NU-560-400
Page 7: ...OM0295 Rev 2 September 2020 Page 7 of 103...
Page 9: ...OM0295 Rev 2 September 2020 Page 9 of 103...
Page 10: ...OM0295 Rev 2 September 2020 Page 10 of 103...
Page 16: ...OM0295 Rev 2 September 2020 Page 16 of 103...
Page 17: ...OM0295 Rev 2 September 2020 Page 17 of 103...
Page 32: ...OM0295 Rev 2 September 2020 Page 32 of 103 Set New Password Retype New Password...
Page 43: ...OM0295 Rev 2 September 2020 Page 43 of 103...
Page 48: ...OM0295 Rev 2 September 2020 Page 48 of 103...
Page 80: ...OM0295 Rev 2 September 2020 Page 80 of 103...
Page 96: ...OM0295 Rev 2 September 2020 Page 96 of 103...
Page 97: ...OM0295 Rev 2 September 2020 Page 97 of 103...
Page 98: ...OM0295 Rev 2 September 2020 Page 98 of 103...
Page 99: ...OM0295 Rev 2 September 2020 Page 99 of 103...
Page 100: ...OM0295 Rev 2 September 2020 Page 100 of 103...
Page 101: ...OM0295 Rev 2 September 2020 Page 101 of 103...