7.11
Digital Airflow Sensor Description & Replacement
7.11.1 Downflow Velocity Sensor
The airflow sensor function utilizes two thermistors that provide a constant current source. One thermistor is a
reference that uses a very low current source. The other thermistor is the airflow measurer that uses a very high
current source. As airflow passes over the thermistors, the airflow removes heat from the thermister measuring
airflow. The loss of heat from the thermistor causes the voltage from the thermistor to increase. This increase
subtracted from the reference thermistor output voltage is what directly relates to airflow velocity. A repeatable
curve can be generated (voltage vs. airflow velocity).
The thermistors used are glass bead and coated and can be cleaned by gently using a cotton swab and alcohol.
Formaldehyde gas, Hydrogen Peroxide and Chlorine Dioxide has no effect on the airflow sensors; however, the
formaldehyde/Ammonium bicarbonate residue that remains after decontamination should be removed from the
airflow sensor thermistors.
Disconnect electrical power from the cabinet before
attempting any maintenance action.
The airflow sensor is removed by turning the locking ring counterclockwise and gently pulling the sensor away
from the connector. To reattach the airflow sensor, turn sensor in keyed connector until key matches, push in and
turn the locking ring clockwise until ring locks. Proceed to calibrate sensor menu.
7.11.2 Exhaust Pressure Sensor
The exhaust pressure sensor is a digital differential velocity pressure flow grid. The exhaust pressure sensor board
is located within the right side of the control center.
The exhaust pressure sensor function utilizes a differential pressure transducer and electronic temperature sensor
to measure exhaust volume. A flow grid mounted in the exhaust airflow stream provides a velocity and static
pressure to the differential pressure transducer. An equation using the pressure drop across the flow grid along
with the temperature reading provides a volumetric flow reading. The volumetric flow reading is then calculated
by the duct area to provide the displayed exhaust volume.
The flow grid located in the exhaust airstream is made from PVC and is not affected by Formaldehyde, Hydrogen
Peroxide or Chlorine Dioxide.
Disconnect electrical power from the cabinet before
attempting any maintenance action.
The exhaust pressure sensor board is removed by unfastening (2) 6/32 nuts and removing the connectors and
tubing then gently pulling up the board until free. To reattach, reverse the above procedure. The exhaust sensor
also has an onboard LED indicator that indicates a properly operating sensor. The LED DS1 blinks in slow 1 second
intervals during normal operation. It will blink faster or full on when a sensor error occurs at which time the
sensor board needs to be replaced. Proceed to calibrate sensor menu.
CAUTION
!
CAUTION
!
Summary of Contents for LabGard NU-560-400
Page 7: ...OM0295 Rev 2 September 2020 Page 7 of 103...
Page 9: ...OM0295 Rev 2 September 2020 Page 9 of 103...
Page 10: ...OM0295 Rev 2 September 2020 Page 10 of 103...
Page 16: ...OM0295 Rev 2 September 2020 Page 16 of 103...
Page 17: ...OM0295 Rev 2 September 2020 Page 17 of 103...
Page 32: ...OM0295 Rev 2 September 2020 Page 32 of 103 Set New Password Retype New Password...
Page 43: ...OM0295 Rev 2 September 2020 Page 43 of 103...
Page 48: ...OM0295 Rev 2 September 2020 Page 48 of 103...
Page 80: ...OM0295 Rev 2 September 2020 Page 80 of 103...
Page 96: ...OM0295 Rev 2 September 2020 Page 96 of 103...
Page 97: ...OM0295 Rev 2 September 2020 Page 97 of 103...
Page 98: ...OM0295 Rev 2 September 2020 Page 98 of 103...
Page 99: ...OM0295 Rev 2 September 2020 Page 99 of 103...
Page 100: ...OM0295 Rev 2 September 2020 Page 100 of 103...
Page 101: ...OM0295 Rev 2 September 2020 Page 101 of 103...