OM0295
Rev 2 September 2020
Page 44 of 103
7.5 Airflow Control System Setup and Calibration
Failure to calibrate airflow to the specified requirements may
result in unsafe conditions of performance (i.e. product and/or
personnel protection, noise and vibration)
7.5.1 General
The operation of the NU-560 cabinet requires that the setup and calibration procedures be performed in order to
certify or commission the cabinet for usage. The setup and calibration procedures performed
ONLY BY THE
CABINET CERTIFIER
ensure that cabinet's setpoints are verified and that the airflow monitor sensors are calibrated
to read the correct values. Press MENU to access Calibration/Service parameter.
Entry into the Calibration/Service functions requires a service password for entry. After pressing the
Calibration/Service menu item, a password screen will appear. The default password is “9876”.
Once the service password is entered, the Calibration/Service menu will appear. To exit, either press the menu or
home icon.
One additional feature for service technicians is to bypass the whole 2 minute warmup time. By pressing the word
warmup, the system will move directly to run. It is recommended that at least a minute of warmup time be
observed, so the airflows can stabilize at their setpoints to avoid an alarm condition.
CAUTION
!
Summary of Contents for LabGard NU-560-400
Page 7: ...OM0295 Rev 2 September 2020 Page 7 of 103...
Page 9: ...OM0295 Rev 2 September 2020 Page 9 of 103...
Page 10: ...OM0295 Rev 2 September 2020 Page 10 of 103...
Page 16: ...OM0295 Rev 2 September 2020 Page 16 of 103...
Page 17: ...OM0295 Rev 2 September 2020 Page 17 of 103...
Page 32: ...OM0295 Rev 2 September 2020 Page 32 of 103 Set New Password Retype New Password...
Page 43: ...OM0295 Rev 2 September 2020 Page 43 of 103...
Page 48: ...OM0295 Rev 2 September 2020 Page 48 of 103...
Page 80: ...OM0295 Rev 2 September 2020 Page 80 of 103...
Page 96: ...OM0295 Rev 2 September 2020 Page 96 of 103...
Page 97: ...OM0295 Rev 2 September 2020 Page 97 of 103...
Page 98: ...OM0295 Rev 2 September 2020 Page 98 of 103...
Page 99: ...OM0295 Rev 2 September 2020 Page 99 of 103...
Page 100: ...OM0295 Rev 2 September 2020 Page 100 of 103...
Page 101: ...OM0295 Rev 2 September 2020 Page 101 of 103...