Instruction manual
OCM Pro CF
®
OCM Pro CF - Rev. 09 as of 07.06.2016
page 137
Fig. 9-126 Error screen
9.6.2 I/O Menu "Digital Inputs ”
Digital input values routed to the transmitter input clamps can be viewed here.
Select from either logically "OFF“ or "ON".
Fig. 9-127 Screen digital values
9.6.3 I/O Menu „Analog Outputs”
Fig. 9-128 Screen analog values
This menu indicates the calculated values which have to be sent to the analog
converter as mA signals. Please note that a transmitter type >S4< allows 4 ana-
log outputs to be set and displayed, analog outputs 1 and 2 however are physi-
cally available only.
Note
The actual current on the output clamps will not be displayed. The only signal
visible is the one the analog output converter is receiving for output purposes.
External faulty wiring neither can be detected nor indicated in this menu.