Instruction manual
OCM Pro CF
page 100
OCM Pro CF - Rev. 09 as of 07.06.2016
®
1 Sensor number (1, 2, 3)
2 Sensor type (wedge, tube, float, Pos-alpha)
3 Installation position (positive, negative)
4 Number of sensors (1x, 2x, 3x)
5 Angle >b< is the angle varying from perpendicular at Pos-Alpha
6 Mounting height of flow velocity sensor
Fig. 9-54
Indication of sensor types
Mounting place
This menu point is to modify the installation height (h) of the flow velocity
sensor. The standard setting is 0 mm. This setting does not need to be modified
unless the wedge or tube sensor has been installed higher or lower. In case of
elevated or depressed installation must be added to or subtracted from the 0 m.
The reference point of the wedge sensor is the bottom edge of the ground plate,
with the pipe sensor it is the horizontal area at the tip of the sensor.
Enter the distance between water surface and the centre of the immersed flow
velocity sensor crystal as soon as the sensor type "Float" has been selected.
If sensor type "float" has been selected, enter the distance from the water
surface to horizontal area at the tip of the pipe sensor here..