Instruction manual
OCM Pro CF
page 74
OCM Pro CF - Rev. 09 as of 07.06.2016
®
1 Maximum Height
2 Maximum Measured Velocity
3 Measurement Window Indication
4 Level Measurement Window no.
5 Level Value
6 Velocity Measurement Window no.
7 Velocity Value
Fig. 9-6
Flow velocity distribution
very good profile, area
free of backwater
good profile, strongly
tending to backwater
suitable profile
ill-suited profile
unsuitable profile
Fig. 9-7
Flow velocity profiles
Note
This function is not available as soon as an OFR radar sensor is connected.
Day totals
The submenu >INFO< (see Fig. 9-8) allows to view the flow totals of the past 7
days (see Fig. 9-9), presumed the transmitter has been operated without any
interruption in the past seven days. Otherwise it shows the total for the
uninterrupted days of operation).
Totalisation normally is carried out at 00:00 h (midnight). If desired, this value can
be modified under RUN – Day Totals - Interval (see Fig. 9-10)
Additionally, you can get information about partial total value since the last reset
(comparable with rout mileage counters in cars). This value can be reset to >0< at
any time by pressing the >ALT<-key. This reset does not influence the totaliser.