Instruction manual
OCM Pro CF
®
OCM Pro CF - Rev. 09 as of 07.06.2016
page 123
CAUTION
Loss of customer settings
Selecting >YES< in Fig. 9-99 (Save new values?) takes the system to the
basic parameter settings. The default settings will be restored and all custom-
er modifications will be set to default condition (system general reset).
System controls set previously will be lost.
When exiting the PAR menu the screen below appears:
Fig. 9-99
Request to complete the system reset
Service code
Additional system setting options are going to be revealed once a special
number has been entered.
These settings are reserved to be used by the NIVUS initial start-up service
since these modifications require comprehensive expert knowledge and do not
need to be adjusted during standard use.
Damping
This menu enables to adjust the display and analog output damping between 20
and 600 seconds. This means that a 0 to 100 % jump of the calculated volume
is going to require the time set to be indicated either on the display as well as on
the output.
Damping, example 1:
Damping 30 seconds, jump from 0 l/s to 100 l/s (=100 %) – instrument needs
30 seconds to run from 0 l/s to 100 l/s.
Damping, example 2:
Damping 30 seconds, jump from 80 l/s to 100 l/s (=20 %) – instrument needs
6 seconds to run from 80 l/s to 100 l/s.
Stability
The period the OCM Pro will operate in without a current level / height meas-
urement value.
The OCM Pro will fall to measurement value >0< according to the damping set
previously if the period expires without a correct level / height measurement
value is being detected.