Instruction manual
OCM Pro CF
page 138
OCM Pro CF - Rev. 09 as of 07.06.2016
®
9.6.4 I/O Menu „Digital Outputs”
Conditions which are calculated by the transmitter and routed to the relay for
output purposes subsequently can be viewed here. Select from either logically
"OFF“ or "ON".
Fig. 9-129 Digital values
Note
The actual output condition of the relay contacts on the output clamps will not
be displayed. The only signal visible is the one the relay is receiving for output
purposes. Mechanical or electrical defects of the output relay cannot be de-
tected this way.
External faulty wiring neither can be detected nor indicated in this menu.
9.6.5 I/O Menu „Sensors”
This menu including the respective submenus allows to view and to assess the
most important sensor conditions. It hence provides information on the quality of
the measurement place, cable layout, echo signal quality and many more pa-
rameters.
In case of using 2 or 3 flow velocity sensors select the sensor by choosing the
according channel number.
Depending on the sensor configuration used (level measurement from bottom
up, level measurement from top down, pressure measurement or measurement
via external sensor) several menus may come up:
Fig. 9-130 Basic selection menu
V-Sensor
Choosing this point is going to bring up a 2-page table including all individual
velocities measured and the heights of the respective measurement windows.