NI 4071 7½-Digit FlexDMM Calibration Procedure
|
© National Instruments
|
91
51. Call the niDMM Read VI.
52. Generate –2.2 A on the calibrator with the current output set to
NORM
.
53. Call the niDMM Read VI.
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_ConfigureMeasurement
Digits
with the following
parameters:
Instrument_Handle
:
The instrument
handle from
niDMM_init
Resolution_Digits
:
6
.
5
Measurement_Function
:
NIDMM_VAL_DC_CURRENT
Range
:
3
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_read
with the
following parameters:
Instrument_Handle
:
The instrument
handle from
niDMM_init
Reading
: The measurement returned
by the function. Verify that this
measurement falls between the limits
listed in Table 25.
Maximum_Time
:
–1
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_read
with the
following parameters:
Instrument_Handle
:
The instrument
handle from
niDMM_init
Reading
: The measurement returned
by the function. Verify that this
measurement falls between the limits
listed in Table 25.
Maximum_Time
:
–1