NI 4071 7½-Digit FlexDMM Calibration Procedure
|
© National Instruments
|
47
23. Generate 45 V at 30 Hz on the calibrator.
Table 6.
niDMM Config Measurement Parameters
Iteration
Calibrator Output
Signal Parameters
niDMM Config Measurement Parameters
Amplitude
Frequency
Function
Range
Resolution
1
4.5 V
50 Hz
NIDMM_VAL_AC_VOLTS
5
6.5
4.5 V
50 Hz
NIDMM_VAL_AC_VOLTS
_DCCOUPLED
5
6.5
2
1 V
1 kHz
NIDMM_VAL_AC_VOLTS
5
6.5
1 V
1 kHz
NIDMM_VAL_AC_VOLTS
_DCCOUPLED
5
6.5
3
2 V
1 kHz
NIDMM_VAL_AC_VOLTS
5
6.5
2 V
1 kHz
NIDMM_VAL_AC_VOLTS
_DCCOUPLED
5
6.5
4
3 V
1 kHz
NIDMM_VAL_AC_VOLTS
5
6.5
3 V
1 kHz
NIDMM_VAL_AC_VOLTS
_DCCOUPLED
5
6.5
5
4.5 V
1 kHz
NIDMM_VAL_AC_VOLTS
5
6.5
4.5 V
1 kHz
NIDMM_VAL_AC_VOLTS
_DCCOUPLED
5
6.5
6
4.5 V
1 kHz
NIDMM_VAL_AC_VOLTS
50
6.5
4.5 V
1 kHz
NIDMM_VAL_AC_VOLTS
_DCCOUPLED
50
6.5
7
4.5 V
20 kHz
NIDMM_VAL_AC_VOLTS
5
6.5
4.5 V
20 kHz
NIDMM_VAL_AC_VOLTS
_DCCOUPLED
5
6.5
8
4.5 V
50 kHz
NIDMM_VAL_AC_VOLTS
5
6.5
4.5 V
50 kHz
NIDMM_VAL_AC_VOLTS
_DCCOUPLED
5
6.5
9
4.5 V
100 kHz
NIDMM_VAL_AC_VOLTS
5
6.5
4.5 V
100 kHz
NIDMM_VAL_AC_VOLTS
_DCCOUPLED
5
6.5
10
4.5 V
300 kHz
NIDMM_VAL_AC_VOLTS
5
6.5
4.5 V
300 kHz
NIDMM_VAL_AC_VOLTS
_DCCOUPLED
5
6.5