NI 4071 7½-Digit FlexDMM Calibration Procedure
|
© National Instruments
|
113
52. Call the niDMM Cal Adjust Offset VI with the following parameters:
53. Call the niDMM Cal Adjust Miscellaneous VI.
54. Call the niDMM Self Cal VI to self-calibrate the NI 4071.
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_CalAdjustOffset
with the following parameters:
Instrument_Handle
:
The instrument
handle from
niDMM_init
Mode
:
NIDMM_VAL_4_WIRE_RES
Range
:
100e3
Input_Resistance
:
NIDMM_VAL_RESISTANCE_NA
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_CalAdjustOffset
with the following parameters:
Instrument_Handle
:
The instrument
handle from
niDMM_init
Mode
:
NIDMM_VAL_4_WIRE_RES
Range
:
10e3
Input_Resistance
:
NIDMM_VAL_RESISTANCE_NA
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_CalAdjustMisc
with
the following parameters:
Instrument_Handle
:
The instrument
handle from
niDMM_init
Type
:
NIDMM_EXTCAL_MISCAL_RREF
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_SelfCal
with the
following parameters:
Instrument_Handle
:
The instrument
handle from
niDMM_init