NI 4071 7½-Digit FlexDMM Calibration Procedure
|
© National Instruments
|
105
10. Call the niDMM Cal Adjust Gain VI with the following parameters:
11. Generate –10 V on the calibrator.
12. Call the niDMM Cal Adjust Gain VI with the following parameters:
13. Disconnect the Fluke 5440 cable from the NI 4071 banana plug connectors, leaving the
other end of the cable connected to the calibrator binding posts.
14. Plug in the insulated banana plug shorting bar across the
HI
and
LO
banana plug connectors
of the NI 4071.
15. Wait two minutes for the thermal EMF to stabilize.
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_CalAdjustGain
with
the following parameters:
Instrument_Handle
:
The instrument
handle from
niDMM_init
Mode
:
NIDMM_VAL_DC_VOLTS
Range
:
10
Input_Resistance
:
NIDMM_VAL_GREATER_THAN_10_
GIGAOHM
Expected_Value
:
10
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_CalAdjustGain
with
the following parameters:
Instrument_Handle
:
The instrument
handle from
niDMM_init
Mode
:
NIDMM_VAL_DC_VOLTS
Range
:
10
Expected_Value
:
–10
Input_Resistance
:
NIDMM_VAL_GREATER_THAN_10_
GIGAOHM