NI 4071 7½-Digit FlexDMM Calibration Procedure
65. Call the niDMM Cal Adjust Offset VI with the following parameters:
66. Call the niDMM Cal Adjust Offset VI with the following parameters:
67. Call the niDMM Cal Adjust Offset VI with the following parameters:
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_CalAdjustOffset
with the following parameters:
Instrument_Handle
:
The instrument
handle from
niDMM_init
Mode
:
NIDMM_VAL_2_WIRE_RES
Range
:
100e3
Input_Resistance
:
NIDMM_VAL_RESISTANCE_NA
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_CalAdjustOffset
with the following parameters:
Instrument_Handle
:
The instrument
handle from
niDMM_init
Mode
:
NIDMM_VAL_2_WIRE_RES
Range
:
10e3
Input_Resistance
:
NIDMM_VAL_RESISTANCE_NA
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_CalAdjustOffset
with the following parameters:
Instrument_Handle
:
The instrument
handle from
niDMM_init
Mode
:
NIDMM_VAL_2_WIRE_RES
Range
:
1e3
Input_Resistance
:
NIDMM_VAL_RESISTANCE_NA