NI 4071 7½-Digit FlexDMM Calibration Procedure
49. Call the niDMM Cal Adjust Offset VI with the following parameters:
50. Call the niDMM Cal Adjust Offset VI with the following parameters:
51. Call the niDMM Cal Adjust Offset VI with the following parameters:
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_CalAdjustOffset
with the following parameters:
Instrument_Handle
:
The instrument
handle from
niDMM_init
Mode
:
NIDMM_VAL_AC_CURRENT
Range
:
1e
–
3
Input_Resistance
:
NIDMM_VAL_RESISTANCE_NA
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_CalAdjustOffset
with the following parameters:
Instrument_Handle
:
The instrument
handle from
niDMM_init
Mode
:
NIDMM_VAL_DC_CURRENT
Range
:
0.01
Input_Resistance
:
NIDMM_VAL_RESISTANCE_NA
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_CalAdjustOffset
with the following parameters:
Instrument_Handle
:
The instrument
handle from
niDMM_init
Mode
:
NIDMM_VAL_AC_CURRENT
Range
:
0.01
Input_Resistance
:
NIDMM_VAL_RESISTANCE_NA