16
|
ni.com
|
NI 4071 7½-Digit FlexDMM Calibration Procedure
24. Call the niDMM Read VI.
25. Call the niDMM Config Measurement VI and select the
Resolution in Digits
instance.
26. Call the niDMM Read VI.
27. Remove the shorting bar from the NI 4071.
28. Reset the calibrator.
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_read
with the
following parameters:
Instrument_Handle
:
The instrument
handle from
niDMM_init
Reading
: The measurement returned
by the function. Verify that this
measurement falls between the limits
listed in Table 20.
Maximum_Time
:
–1
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_ConfigureMeasurement
Digits
with the following
parameters:
Instrument_Handle
:
The instrument
handle from
niDMM_init
Resolution_Digits
:
7.5
Measurement_Function
:
NIDMM_VAL_DC_VOLTS
Range
:
1000
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_read
with the following
parameters:
Instrument_Handle
:
The instrument
handle from
niDMM_init
Reading
: The measurement returned
by the function. Verify that this
measurement falls between the limits
listed in Table 20.
Maximum_Time
:
–1