NI 4071 7½-Digit FlexDMM Calibration Procedure
|
© National Instruments
|
65
14. Call the niDMM Read VI. Verify that this measurement falls between the tolerances listed
in Table 24.
15. Call the niDMM Config Measurement VI and select the
Resolution in Digits
instance.
16. Call the niDMM Configure Offset Comp Ohms VI with
Offset Compensated Ohms
set to
NIDMM_VAL_OFFSET_COMP_OHMS_ON
.
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_read
with the
following parameters:
Instrument_Handle
:
The instrument
handle from
niDMM_init
Reading
: The measurement returned
by the function. Verify that this
measurement falls between the limits
listed in Table 24.
Maximum_Time
:
–1
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_ConfigureMeasurement
Digits
with the following
parameters:
Instrument_Handle
:
The instrument
handle from
niDMM_init
Resolution_Digits
:
7
.
5
Measurement_Function
:
NIDMM_VAL_2_WIRE_RES
Range
:
1e3
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_ConfigureOffsetComp
Ohms
with the following parameters:
Instrument_Handle
:
The instrument
handle from
niDMM_init
OffsetCompOhms
:
NIDMM_VAL_OFFSET_COMP_OHMS
_ON