114
|
ni.com
|
NI 4071 7½-Digit FlexDMM Calibration Procedure
55. Generate 0
on the calibrator with external sense turned on, but with 2-wire compensation
turned off.
56. Call the niDMM Cal Adjust Offset VI with the following parameters:
57. Call the niDMM Cal Adjust Offset VI with the following parameters:
58. Call the niDMM Cal Adjust Offset VI with the following parameters:
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_CalAdjustOffset
with the following parameters:
Instrument_Handle
:
The instrument
handle from
niDMM_init
Mode
:
NIDMM_VAL_4_WIRE_RES
Range
:
10e6
Input_Resistance
:
NIDMM_VAL_RESISTANCE_NA
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_CalAdjustOffset
with the following parameters:
Instrument_Handle
:
The instrument
handle from
niDMM_init
Mode
:
NIDMM_VAL_4_WIRE_RES
Range
:
1e6
Input_Resistance
:
NIDMM_VAL_RESISTANCE_NA
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_CalAdjustOffset
with the following parameters:
Instrument_Handle
:
The instrument
handle from
niDMM_init
Mode
:
NIDMM_VAL_4_WIRE_RES
Range
:
1e3
Input_Resistance
:
NIDMM_VAL_RESISTANCE_NA