NI 4071 7½-Digit FlexDMM Calibration Procedure
6.
Repeat step 5 for each of the remaining iterations listed in Table 10.
You have completed verifying the 4-wire resistance of the NI 4071. Select one of the following
options:
•
If you want to continue verifying other modes, go to the
section.
•
If you do
not
want to verify other modes
and
you are performing a
post-adjustment
verification, go to the
Completing the Adjustment Procedures
section.
•
If you do
not
want to verify any additional modes
and
you are performing a
pre-adjustment
verification, call the niDMM Close VI to close the session.
Table 10.
niDMM Config Measurement Parameters
Iteration
Calibrat
or
Output
niDMM Config Measurement Parameters
OffsetComp
Ohms
Function
Range
Resolution
1
10 M
NIDMM_VAL_4_
WIRE_RES
10e6
7.5
OFF
2
1 M
NIDMM_VAL_4_
WIRE_RES
1e6
7.5
OFF
3
100 k
NIDMM_VAL_4_
WIRE_RES
100e3
7.5
OFF
4
10 k
NIDMM_VAL_4_
WIRE_RES
10e3
7.5
ON
5
1 k
NIDMM_VAL_4_
WIRE_RES
1e3
7.5
ON
6
100
NIDMM_VAL_4_
WIRE_RES
100
7.5
ON
7
0
NIDMM_VAL_4_
WIRE_RES
10e6
7.5
OFF
8
0
NIDMM_VAL_4_
WIRE_RES
1e6
7.5
OFF
9
0
NIDMM_VAL_4_
WIRE_RES
100e3
7.5
OFF
10
0
NIDMM_VAL_4_
WIRE_RES
10e3
7.5
ON
11
0
NIDMM_VAL_4_
WIRE_RES
1e3
7.5
ON
12
0
NIDMM_VAL_4_
WIRE_RES
100
7.5
ON