NI 4071 7½-Digit FlexDMM Calibration Procedure
|
© National Instruments
|
59
d.
Call the niDMM Read VI. Verify that this measurement falls between the tolerances
listed in Table 23. Tolerances are provided instead of absolute limits, because your
calibrator will have different discrete resistance values.
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_read
with the
following parameters:
Instrument_Handle
:
The instrument
handle from
niDMM_init
Reading
: The measurement returned
by the function. Verify that this
measurement falls between the limits
listed in Table 23.
Maximum_Time
:
–1