NI 4071 7½-Digit FlexDMM Calibration Procedure
|
© National Instruments
|
107
23. Generate 10 M
from the calibrator without external sense.
24. Call the niDMM Cal Adjust Gain VI with the following parameters:
25. Generate 1.9 M
from the calibrator without external sense.
26. Call the niDMM Cal Adjust Gain VI with the following parameters:
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_CalAdjustGain
with
the following parameters:
Instrument_Handle
:
The instrument
handle from
niDMM_init
Mode
:
NIDMM_VAL_2_WIRE_RES
Range
:
10e6
Input_Resistance
:
NIDMM_VAL_RESISTANCE_NA
Expected_Value
:
the display on the
calibrator for 0 W
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_CalAdjustGain
with
the following parameters:
Instrument_Handle
:
The instrument
handle from
niDMM_init
Mode
:
NIDMM_VAL_2_WIRE_RES
Range
:
10e6
Input_Resistance
:
NIDMM_VAL_RESISTANCE_NA
Expected_Value
:
the display on the
calibrator for 10 MW
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_CalAdjustGain
with
the following parameters:
Instrument_Handle
:
The instrument
handle from
niDMM_init
Mode
:
NIDMM_VAL_2_WIRE_RES
Range
:
10e6
Input_Resistance
:
NIDMM_VAL_RESISTANCE_NA
Expected_Value
:
the display on the
calibrator for 1.9 MW