NI 4071 7½-Digit FlexDMM Calibration Procedure
|
© National Instruments
|
109
34. Call the niDMM Cal Adjust Gain VI with the following parameters:
35. Generate 0
from the calibrator without external sense or 2-wire compensation.
36. Call the niDMM Cal Adjust Gain VI with the following parameters:
37. Call the niDMM Cal Adjust Offset VI with the following parameters:
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_CalAdjustGain
with
the following parameters:
Instrument_Handle
:
The instrument
handle from
niDMM_init
Mode
:
NIDMM_VAL_2_WIRE_RES
Range
:
100e6
Input_Resistance
:
NIDMM_VAL_RESISTANCE_NA
Expected_Value
: The value that is
displayed on the calibrator for
100 M
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_CalAdjustGain
with
the following parameters:
Instrument_Handle
:
The instrument
handle from
niDMM_init
Mode
:
NIDMM_VAL_2_WIRE_RES
Range
:
100e6
Input_Resistance
:
NIDMM_VAL_RESISTANCE_NA
Expected_Value
:
The value that is
displayed on the calibrator for 0
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_CalAdjustOffset
with the following parameters:
Instrument_Handle
:
The instrument
handle from
niDMM_init
Mode
:
NIDMM_VAL_2_WIRE_RES
Range
:
100e6
Input_Resistance
:
NIDMM_VAL_RESISTANCE_NA