NI 4071 7½-Digit FlexDMM Calibration Procedure
|
© National Instruments
|
53
31. Refer to Table 8 for the appropriate calibrator output signal and measurement parameter
values as you complete the following steps:
a.
On the calibrator, generate the value listed under Calibrator Output in Table 8 for the
current iteration.
b.
Call the niDMM Config Measurement VI with
Function
set to
NIDMM_VAL_AC_VOLTS
, and set the remaining parameters as shown in Table 8 for
the current iteration. Select the
Resolution in Digits
instance of the
niDMM Config
Measurement VI.
c.
Call the niDMM Read VI.
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_ConfigureMeasurement
Digits
with the following
parameters:
Instrument_Handle
:
The instrument
handle from
niDMM_init
Resolution_Digits
: Set as shown in
Table 8 for the current iteration
Measurement_Function
:
NIDMM_VAL_AC_VOLTS
Range
: Set as shown in Table 8 for
the current iteration
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_read
with the
following parameters:
Instrument_Handle
:
The instrument
handle from
niDMM_init
Reading
: The measurement returned
by the function. Verify that this
measurement falls between the limits
listed in Table 21.
Maximum_Time
:
–1