NI 4071 7½-Digit FlexDMM Calibration Procedure
|
© National Instruments
|
11
12. Call the niDMM Read VI.
13. Call the niDMM Config Measurement VI and select the
Resolution in Digits
instance.
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_read
with the
following parameters:
Instrument_Handle
:
The instrument
handle from
niDMM_init
Reading
: The measurement returned
by the function. Subtract the
previously stored 100 mV 10 M
mode offset value from
this measurement.Verify that this
measurement falls between the limits
listed in Table 20.
Maximum_Time
:
–1
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_ConfigureMeasurement
Digits
with the following
parameters:
Instrument_Handle
:
The instrument
handle from
niDMM_init
Resolution_Digits
:
7.5
Measurement_Function
:
NIDMM_VAL_DC_VOLTS
Range
:
1