NI 4071 7½-Digit FlexDMM Calibration Procedure
22. Call the niDMM Read VI.
23. Generate –100
A on the calibrator with the current output set to
NORM
.
24. Call the niDMM Read VI.
25. Replace the shunt recommended for 100
A currents with the shunt recommended for
1 mA as indicated in Table 13.
26. Generate 1 mA on the calibrator with the current output set to
NORM
.
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_read
with the
following parameters:
Instrument_Handle
:
The instrument
handle from
niDMM_init
Reading
: The measurement returned
by the function. Verify that this
measurement falls between the limits
listed in Table 25.
Maximum_Time
:
–1
LabVIEW Block Diagram
C/C++ Function Call
Call
niDMM_read
with the
following parameters:
Instrument_Handle
:
The instrument
handle from
niDMM_init
Reading
: The measurement returned
by the function. Verify that the
reading or measurement value falls
between the limits listed in Table 25.
Maximum_Time
:
–1